Generate Catalog For: Gas Purification Glove Box ADPGB-504

Gas Purification Glove Box ADPGB-504 features a single-station design with a 1500 × 760 × 900 mm main chamber for safe and efficient sample handling. It is equipped with anodized aluminium oxide glove ports and O-ring seals to ensure leak-proof, secure operation. A built-in automatic vacuum pump start-stop function enhances operational efficiency. Integrated LED lighting provides a bright workspace for precise handling. Our glove box is widely used in pharmaceutical, biotechnology, and clean energy research applications.

Features :
  • Stainless Steel Body
  • Butyl Rubber Gloves
  • Low H₂O and O₂
  • PLC Pressure Control
  • Touch Screen Display
Specifications :
Working Gas Nitrogen, argon (gas purity 99.999%)
H₂O and O₂ Index Less than 1PPM
Circulating Capacity Integrated fan flow rate of 90m³/h
Material Stainless steel, 3mm thickness
Main Chamber Dimension (L×B×H) 1500 × 760 × 900mm
Antechamber Dimension(D×L) Large:377×600, Small:150×300
Sight Window Inclined, tempered glass, 8mm thickness
Number of Glove Port 3
Glove Port Material Anodized aluminum oxide, O-ring sealed
Gloves Butyl rubber, diameter 8-inch, length 32 inch
Leakage Rate Less than 0.001 vol%/h
Pressure Control PLC controlled, ±10 mbar adjustable, ±12 mbar protection
Filter 0.3 microns, Qty 2 (gas inlet and outlet)
Shelving 2 pieces
Purification Material Copper catalyst: 5kg, Molecular sieve: 7kg
Purification Capacity Deoxygenation: 60L, Dehydration: 2kg
Lighting LED lighting
Regeneration Gas Mixture of working gas and 5% hydrogen gas
Vacuum Pump 8m³/h, manual or PLC start
Vacuum Degree 5x10-2Pa
Vacuum Pressure Gauge Mechanical dial, measures transition chamber vacuum
Main Valve DN 40 KF, pneumatic vacuum flapper valve
Control Valve Solenoid integrated valves
Control System Siemens PLC touch screen, self-diagnosis, auto start, cycle & password protection
Display System Displays status, pressure, records
Water Analyzer Range 0 to 1000 ppm
Oxygen Analyzer Range 0 to 1000 ppm, Accuracy ±0.1 ppm
Foot Switch Controls gas replenishment/exhaust, pressure fine-tuning
Total Dimension (L×B×H) 2240 × 790 × 1860mm
Weight 850kg
Applications :

Gas Purification Glove Box ADPGB-504 is used in labs for chemical, biotech, and material research applications. It maintains precise environmental control while providing clear observation.