Generate Catalog For: Double Beam Spectrophotometer ADDS-524

Double Beam UV-Vis Spectrophotometer ADDS-524 analyses samples in the visible to near-infrared range from 190 to 1100 nm. The sample compartment supports various cell holders for flexible handling. It provides real-time graphical display of spectra and kinetic curves for better visualization. This unit includes error alerts to ensure optimal performance. Our Double Beam UV-Vis Spectrophotometer is used in pharmaceuticals, biotechnology, and chemical industries for accurate analysis of absorbance. 

Features :
  • Data storage and transfer
  • High photometric accuracy
  • User-friendly interface
  • Improved baseline stability
  • Automatic wavelength scanning
Specifications :
Optical System Double Beam, Grating 1200 lines/mm
Wavelength Range 190 to 1100 nm
Spectral Bandwidth 2 nm
Wavelength Setting Auto
Wavelength Accuracy ± 0.1 nm 656.1 nm, ± 0.3 nm all
Wavelength Repeatability <= 0.1 nm
Photometric Accuracy 0.2 % T (0 to 100 % T), ± 0.002 A (0to 0.5 A), ± 0.004 A (0.5 to 1 A)
Photometric Repeatability ≤ 0.15 % T (0 to 100 % T), 0.001 A (0 to 0.5 A), 0.002 A (0.5 to 1 A)
Photometric Stability ± 0.15 % T, ± 0.002 A (0 to 0.5 A), ± 0.004 A (0.5 to 1 A)
Photometric Range 0 to 200 % T, -0.3 to 3 A, 0 to 9999 C
Stray Light <= 0.03 % T 220 nm, 360 nm
Stability ± 0.0005 A/h 500 nm
Work Mode T, A, C, E
Baseline Flatness ± 0.002 A
Noise 0.0005 A 500 nm
Scanning Speed Hi, Med, Low (Max. 3000nm/min)
Display 7-inch Colour touch screen
Light Source Deuterium and tungsten lamp
Detector Imported silicon photodiode
Cuvette Holder 10mm single hole cell holder
Output USB drive, USB host, RS232
Power AC 110 to 220 V 50 to 60 Hz
Dimensions (L×B×H) 810×660×390 mm
Gross Weight 27 Kg
Applications :

Our Double Beam Spectrophotometer ADDS-524 is used in clinical diagnostics, academic research, and water quality monitoring, for accurate chemical and optical measurements.