Generate Catalog For: Single Beam UV-Vis Spectrophotometer ADSS-512

Single Beam UV-Vis Spectrophotometer ADSS-512 comes with ± 0.1 nm wavelength accuracy for ensuring precise detection of absorbance peaks. It is compatible with various cuvette types, offering flexibility for different sample types. The compact design makes it ideal for laboratory environments with limited space. This unit features automatic baseline correction to ensure stable and precise readings. Our Single Beam UV-Vis Spectrophotometer is used in chemical research, food testing, and academic laboratories for measurement of absorbance.

Features :
  • Fast scan capabilityData storage optionsVersatile cuvette holderHigh accuracyWavelength range
Specifications :
Optical System Single beam, grating 1200 lines/mm
Wavelength Range 190 to 1100 nm
Spectral Bandwidth 0.5/1/2/4/5nm
Wavelength Accuracy ± 0.1nm 656.1nm, ± 0.3nm all
Wavelength Repeatability <=0.1nm
Photometric Accuracy 0.2 % T (0 to 100 % T), ± 0.002A (0 to 0.5 A), ± 0.004 A (0.5 to 1 A)
Photometric Repeatability ≤0.15 % T (0 to 100 % T), 0.00 A (0 to 0.5A), 0.002 A (0.5 to 1A)
Photometric Range 0 to 200 % T, 0.3 to 3A, 0 to 999 C
Stray Light <=0.03 % T, 220 nm, 360 nm
Stability ± 0.0005A/h 500 nm
Baseline Flatness ± 0.001A
Noise 0.0005A 500 nm
Work Mode T, A, C, E
Scanning Speed High, medium, low (Max. 3000 nm/min)
Display 10.1 inches Colour touch screen
Light Source Deuterium and tungsten lamp
Detector Imported silicon photodiode
Cuvette Holder 10 mm 4-Cell holder
Output USB drive, USB host, RS232
Power AC 110-220V 50-60Hz
Dimensions (L×B×H) 810×660×390 mm
Gross Weight 27 kg
Applications :

Our Single Beam UV-Vis Spectrophotometer ADSS-512 is used in chemical analysis, food quality control, and forensics for accurate measurement of absorbance and substance identification.