Single Beam UV-Vis Spectrophotometer ADSS-510 analyses samples in the visible to near-infrared range from 190 to 1100 nm. The compact design makes it ideal for routine analytical laboratories. It uses a single optical path, allowing direct measurement of absorbance for a single sample at a time. This unit supports multiple cuvette types and path lengths for flexible sample analysis. Our Single Beam UV-Vis Spectrophotometer is used in chemistry, food analysis, and life sciences for quantitative analysis of a sample based on light absorption.
| Optical System | Single beam, grating 1200 lines/mm |
|---|---|
| Wavelength Range | 190 to 1100 nm |
| Spectral Bandwidth | 2 nm |
| Wavelength Accuracy | ± 0.1nm 656.1nm, ± 0.3nm all |
| Wavelength Repeatability | <=0.1nm |
| Photometric Accuracy | 0.2 % T (0 to 100 % T), ± 0.002A (0 to 0.5 A), ± 0.004 A (0.5 to 1 A) |
| Photometric Repeatability | ≤0.15 % T (0 to 100 % T), 0.00 A (0 to 0.5A), 0.002 A (0.5 to 1A) |
| Photometric Range | 0 to 200 % T, 0.3 to 3A, 0 to 999 C |
| Stray Light | <=0.03 % T, 220 nm, 360 nm |
| Stability | ± 0.0005A/h 500nm |
| Baseline Flatness | ± 0.002 A |
| Noise | 0.0005 A 500nm |
| Work Mode | T, A, C, E |
| Scanning Speed | High, medium, low (Max. 3000 nm/min) |
| Display | 10.1 inches Colour touch screen |
| Light Source | Deuterium and tungsten lamp |
| Detector | Imported silicon photodiode |
| Cuvette Holder | 10 mm 4-Cell holder |
| Output | USB drive, USB host, RS232 |
| Power | AC 110-220V 50-60Hz |
| Dimensions (L×B×H) | 810×660×390 mm |
| Gross Weight | 27 kg |
Our Single Beam UV-Vis Spectrophotometer ADSS-510 is used in water and wastewater analysis, forensic science, and biotechnology research for precise quantitative and qualitative analysis of samples.