| Sample Volume | 1 to 100µl, 0.1µl / step |
| Reagent Position | 2 × 40 |
| Reagent Volume R1 | 1 to 400 µl, 1µl / step |
| Reagent Volume R2 | 1 to 400 µl, 1µl / step |
| Wavelength | 340 nm to 810 nm |
| Throughput 1 Reagent | 300 tests/hour |
| Throughput 2 Reagent | 200 tests/hour |
| Reaction Time | 0 to 999 seconds |
| Assay Items | 40 Assay |
| Operation | Random access |
| Method |
End Point Kinetic Fixed Time Immunoturbidimetric 1 to 2 reagents Multistandard Reagent / serum blank |
| Reagent / Sample Probe |
With liquid level sensor Collision sensor Teflon coating Automatic washing plastic |
| Washing | 8 Channel automatic washing system for reaction cuvette |
| Calibration |
Linear Non- linear Multi-standard K factor |
| Sample Dilution / Retest | Sample can be diluted and rested if results are out of range or sample is insufficient. Also, with calibrator auto diluting function. |
| Light Source | Halogen Lamp |
| Absorbance Resolution | 0.0001 A |
| Absorbance Accuracy | ± 0.0003A (0 to 2.5 A) |
| Repeatability (CV %) | ≤ 2 % |
| Quality Control | Levy Jennings quality control Program at 3 levels |
| Dimensions (L× B × H) | 850 × 720 × 530 mm |
| Weight | 90 kg |
Automatic Biochemistry AnalyzerADBCA-502 quantifies biochemical substances in biological samples, aiding accurate diagnostics and monitoring. It plays a vital role in treatment planning and advancing medical research.
1. What is random access mode in Automatic Biochemistry Analyzer ADBCA-502, and how does it benefit laboratory Operations?
Automatic Biochemistry Analyzer ADBCA-502 operates in random access mode, allowing multiple tests on different samples simultaneously. This non-sequential processing boosts flexibility and speed in lab operations. It enables on-demand testing without waiting for batch completion. Ideal for high-volume clinical labs requiring efficient and precise workflows.
2. What are the different types of calibration methods used in Automatic Biochemistry Analyzer ADBCA-502, and how do they work?
Automatic Biochemistry Analyzer ADBCA-502 uses various calibration methods for accuracy, they are linear, Non-linear, Multi-standard, and K factor calibration. Linear calibration assumes a straight-line relationship between signal and concentration whereas Non-linear calibration applies curves. Multi-standard uses several standards, while K factor uses a fixed constant for quick results.
3. What does repeatability mean in the Automatic Biochemistry Analyzer ADBCA-502, and why is it crucial?
Automatic Biochemistry Analyzer ADBCA-502 ensures high repeatability ?2% for consistent and reliable result. Repeatability is the Analyzer’s consistency in testing the same sample multiple times. Measured by CV%, a value of ? 2% indicates high reliability. High repeatability is essential for accurate diagnoses and patient monitoring. It ensures trustworthy results in clinical labs.
4. What is the throughput (tests per hour) of the Automatic Biochemistry Analyzer ADBCA-502 using one or two reagents?
Throughput is the number of tests in Automatic Biochemistry Analyzer ADBCA-502 completes in one hour, impacting lab efficiency. This Analyzer can perform up to 200 single-reagent tests per hour. For two-reagent tests, throughput is up to 150 tests per hour. This accounts for extra time needed for mixing and processing.
5. Does Automatic Biochemistry Analyzer ADBCA-502 have safety and automation features?
Automatic Biochemistry Analyzer ADBCA-502 includes liquid level detection to prevent sampling errors. It features anti-collision sensors that stop the probe if an obstruction is detected. Automatic interior and exterior probe washing ensures contamination-free analysis. Real-time alarms alert users to issues like reagent shortages or full waste. Self-diagnostic and auto-calibration functions enhance safety and reduce manual work.
6. What are Automatic Biochemistry Analyzers?
Automatic Biochemistry Analyzers are fully automated laboratory instrument used to measure chemical substances in biological samples. They reduce manual effort and increase accuracy and efficiency in clinical diagnostics. Designed to support a wide range of biochemical tests essential for disease diagnosis and monitoring. Such Analyzers is commonly used in hospitals, diagnostic labs, and research institutions.
7. What are the advanced features that make Automatic Biochemistry Analyzers efficient and reliable in laboratory settings?
Automatic Biochemistry Analyzers offer reagent temperature regulation to ensure stable reactions. They include liquid level detection for precise sampling and collision safety to protect internal components. Analyzers have high-performance cuvette washer that maintains cleanliness and prevents cross-contamination. These features collectively enhance testing accuracy, safety, and operational efficiency.
8. What makes Automatic Biochemistry Analyzers enhanced?
Automatic Biochemistry Analyzers feature full automation, reducing manual steps and improving accuracy. They have advanced software platforms enable efficient data management and LIS integration. Eco-friendly features like waste reduction and sustainable reagent use improve lab sustainability. Overall, these Analyzers deliver faster, safer, and more reliable diagnostic results in high-throughput environments.
| Sample Volume | 1 to 100µl, 0.1µl / step |
| Reagent Position | 2 × 36 |
| Reagent Volume R1 | 1 to 400 µl, 1µl / step |
| Sample Volume | 1 to 100µl, 0.1µl / step |
| Reagent Position | 2 × 40 |
| Reagent Volume R1 | 1 to 400 µl, 1µl / step |
| Sample Type: | Serum, whole Blood, Plasma |
| Sample Volume: | 45 µl |
| Detection Time: | ≤3 minutes |