Heated Lid Dry Bath Incubator ADHDB-505
Heated Lid Dry Bath Incubator ADHDB-505
Heated Lid Dry Bath Incubator ADHDB-505

Heated Lid Dry Bath Incubator ADHDB-505

Heated Lid Dry Bath Incubator ADHDB-505 allows temperature settings within the range of -10℃ to 100℃. It consists of modular heating as well as cooling blocks designed for efficient temperature control. This system has LCD display that provides real-time operation information for easy monitoring. Designed with interchangeable blocks, this unit adapts easily to different tube sizes. Our Heated Lid Dry Bath Incubator is used for precise thermal control of biological or chemical samples in laboratory incubation processes.


Temperature Setting Range -10℃ to 100℃
Temperature Control Range RT -25℃ to 100℃
Temperature Accuracy ≤ ± 0.3℃ at 37℃
Temperature Uniformity ≤ ± 0.3℃ at 37℃
Display Accuracy 0.1℃
Heating Time ≤ 12min (from 25℃ to 100℃)
Cooling Time ≤ 12min (from 100℃ to 25℃)
≤ 25min (from RT. to RT.-25℃)
Heating Lid Temperature Setting Range OFF to +10℃
Heating Lid Temperature Accuracy ≤ ± 0.5℃
Heating Lid Heating Time ≤ 10min (from 25℃ to 100℃)
Multi-point Running Yes (maximum 5 points)
Multi Cycle Yes (maximum 99 times)
Auto Running Yes
Auto Resume to Run Yes
Auto Heating Yes
Power Requirements AC 220V or 100V, 50 to 60 Hz, 150 W
Fuse 250V 3A Ф 5×20
Dimension (W × D × H) 300 x 200 x 210 mm
Net Weight 4.7 kg

Heated Lid Dry Bath Incubator ADHDB-505 is used for controlled heating or cooling of biological and chemical samples in research and clinical laboratories. It is ideal for applications such as enzyme reactions, sample incubation, and nucleic acid denaturation.

Block No.

Description

 

Temperature RangeQuantity
Block 01 (Heating and Cooling)

35 × 2.0 ml centrifuge tube

 

RT -25℃ to 100℃1 Pc
Block 02 (Heating and Cooling)

35 × 1.5 ml centrifuge tube

 

RT -25℃ to 100℃1 Pc
Block 03 (Heating and Cooling)

54 × 0.5 ml centrifuge tube

 

RT -25℃ to 100℃1 Pc
Block 04 (Heating and Cooling)

96 × 2.0 ml PCR centrifuge tube

 

RT -25℃ to 100℃ 1 Pc
Block 05 (Heating and Cooling)

96 × 0.2 ml ELISA plate

 

RT -25℃ to 100℃1 Pc
Block 06 (Heating and Cooling)

35 × 2.0 ml centrifuge tube

 

RT +5℃ to 100℃1 Pc
Block 07 (Heating and Cooling)

35 × 1.5 ml centrifuge tube

 

RT +5℃ to 100℃1 Pc
Block 08 (Heating and Cooling)

54 × 0.5 ml centrifuge tube

 

RT +5℃ to 100℃1 Pc
Block 09 (Heating and Cooling)

96 × 0.2 ml PCR centrifuge tube

 

RT +5℃ to 100℃1 Pc
Block 10 (Heating and Cooling)

96 × 0.2 ml ELISA plate

 

RT +5℃ to 100℃1 Pc

Frequently Asked Questions of Heated Lid Dry Bath Incubator ADHDB-505

1. What types of samples can the Heated Lid Dry Bath Incubator ADHDB-505 can accommodate?

Heated Lid Dry Bath Incubator ADHDB-505 is designed to accommodate a variety of sample types commonly used in laboratories. It can hold centrifuge tubes of multiple sizes, including small and large volumes, as well as 96-well microplates such as ELISA plates. This versatility makes them suitable for a wide range of biological and chemical samples.

2. What temperature range does the Heated Lid Dry Bath Incubator ADHDB-505 support?

Heated Lid Dry Bath Incubator ADHDB-505 support a broad temperature range that covers both heating and cooling needs in laboratory settings. It can maintain precise temperature control from below room temperature up to levels suitable for most incubation and reaction processes. This flexibility allows researchers to carry out various sensitive assays and sample preparations. The reliable temperature stability ensures consistent results across different experimental conditions.

3. How easy it is for Heated Lid Dry Bath Incubator ADHDB-505 to switch between different blocks?

In Heated Lid Dry Bath Incubator ADHDB-505 witching between different blocks is designed to be quick and straightforward. The blocks are engineered for easy insertion and removal without the need for special tools, allowing users to change sample formats efficiently. This flexibility supports diverse laboratory workflows and reduces downtime between experiments. As a result, users can adapt the incubator to different applications with minimal effort.

4. How is temperature accuracy maintained in the Heated Lid Dry Bath Incubator ADHDB-505?

Temperature accuracy in the Heated Lid Dry Bath Incubator ADHDB-505 is maintained through the use of precise temperature sensors combined with advanced PID control technology. This system continuously monitors and adjusts the heating or cooling to ensure stable and uniform temperature distribution. The design of the blocks also promotes even heat transfer to all samples, minimizing temperature variation.

5. What are the typical applications of the Heated Lid Dry Bath Incubator ADHDB-505?

Heated Lid Dry Bath Incubator ADHDB-505 is commonly used for temperature-sensitive laboratory procedures such as enzyme reactions, nucleic acid denaturation, and sample incubation. It provide precise heating and cooling control essential for biochemical assays and molecular biology experiments. The unit’s versatility makes it suitable for clinical diagnostics, research, and pharmaceutical development. This ensures reliable results across a wide range of scientific applications.

6. What is a Heated Lid Dry Bath Incubator?

A Heated Lid Dry Bath Incubator is a laboratory instrument designed to maintain precise temperatures for samples in vials, tubes, or other containers. It uses a heated lid to ensure consistent, uniform temperature distribution. Unlike water-based baths, this type of incubator operates without liquids, preventing contamination and the risk of spills.

7. How does the Heated Lid Dry Bath Incubator work?

The Heated Lid Dry Bath Incubator works by using a heating system within the lid and sometimes the base. The temperature is controlled to provide a consistent environment for samples. The heated lid ensures that the top portion of the samples remains heated, reducing the risk of condensation or temperature gradients. This dual-heating system ensures uniform temperature distribution, improving the accuracy and reliable results.

8. What are the advantages of using a Heated Lid Dry Bath Incubator?

A Heated Lid Dry Bath Incubator offers several advantages over traditional water baths or other incubation methods. One of the main benefits is the absence of liquid, which eliminates the risk of contamination, spillage, and the need for frequent cleaning associated with water baths. The heated lid ensures uniform temperature distribution, reducing temperature gradients and minimizing condensation.

9. Can I use a Heated Lid Dry Bath Incubator for different types of samples?

Yes, Heated Lid Dry Bath Incubator are versatile and can be used for a wide range of sample types, including DNA/RNA extraction, enzyme reactions, cell culture, and more. The incubator typically supports various block types to accommodate different test tubes and microplates.

10. How to clean Heated Lid Dry Bath Incubator?

To clean a Heated Lid Dry Bath Incubator, first ensure the unit is turned off and completely cooled down. Wipe the exterior and blocks with a soft cloth dampened with mild soap and water. Avoid harsh chemicals and abrasives. For stains, use a gentle disinfectant. Ensure no liquid enters the interior and let it dry before use.

Similar Products of Heated Lid Dry Bath Incubator

Temperature Range RT+5℃ to 100℃
Temperature Accuracy ±0.3℃
Temperature Set Range 0℃ to100℃
Block Temperature Setting Range RT -30°C to 105°C
Temperature Set Range -10°C to 105°C
Lid Temperature Setting Range RT +5°C to 110°C
Temperature Setting Range 0℃ to 100℃
Temperature Control Range RT +5℃ to 100℃
Temperature Accuracy ≤ ± 0.3℃ at 37℃
Block Temperature Setting Range RT +5℃ to 105℃
Temperature Set Range 0℃ to 105℃
Lid Temperature Setting Range RT +5°C to 110°C

Dry Bath Incubator

Dry Bath Incubator ADDB-514
Temperature Range:RT + 5℃ to 150℃
Heating Up Rate:≤ 10 min, 20℃ to 100℃ , ≤ 30 min, 20℃ to 150℃
Temperature Stability :±0.3℃ at 40℃ , ±0.5℃ at 100℃ , ±1℃ at 120℃
Dry Bath Incubator ADDB-501
Temperature Setting Range:0℃ to 100℃
Temperature Control Range:RT +5℃ to 100℃
Temperature Accuracy:≤ ± 0.5℃ at 37℃
Dry Bath Incubator ADDB-507
Temperature Range:RT+5℃ to 105℃
Temperature Accuracy:±0.3℃
Temperature Set Range:0℃ to 105℃
Dry Bath Incubator ADDB-502
Temperature Setting Range:0℃ to 100℃
Temperature Control Range:RT+5℃ to 100℃
Temperature Accuracy:≤ ± 0.3℃
Dry Bath Incubator ADDB-508
Temperature Range:RT+5°C to 100°C
Temperature Accuracy:±0.3°C
Temperature Set Range:5°C to 100°C
Dry Bath Incubator ADDB-510
Temperature Range:RT + 5℃ to 150℃
Heating Up Rate:≤ 10 min, 20℃ to 100℃ , ≤ 30 min, 20℃ to 150℃
Temperature Stability :±0.3℃ at 40℃ , ±0.5℃ at 100℃ , ±1℃ at 120℃
Dry Bath Incubator ADDB-518
Temperature Range:RT. 5℃ to 130℃
Temperature Accuracy:± 0.5℃ at 40℃ , ± 1℃ at 40℃
Temperature Uniformity:± 0.5℃
Dry Bath Incubator ADDB-503
Temperature Setting Range:0℃ to 100℃
Temperature Control Range:RT. -20℃ to 100℃
Temperature Accuracy:≤ ± 0.3℃
Dry Bath Incubator ADDB-504
Temperature Setting Range:0℃ to 100℃
Temperature Control Range:RT 5℃ to 100℃
Temperature Accuracy:≤±0.3℃
Dry Bath Incubator ADDB-506
Temperature Setting Range:0℃ to 110℃
Temperature Control Range:RT +5℃ to 110℃
Temperature Accuracy:≤ ± 0.3 at 37℃
Dry Bath Incubator ADDB-517
Temperature Control Range:RT. +5℃ to 160℃
Temperature Accuracy:± 0.5℃ at 40℃ , ± 1℃ at 120℃
Temperature Uniformity:± 0.5℃
Dry Bath Incubator ADDB-515
Temperature Range:RT + 5℃ to 150℃
Heating Up Rate:≤ 10 min, 20℃ to 100℃ , ≤ 30 min, 20℃ to 150℃
Temperature Stability :±0.3℃ at 40℃ , ±0.5℃ at 100℃ , ±1℃ at 120℃
Dry Bath Incubator ADDB-505
Temperature Setting Range:-10℃ to 100℃
Temperature Control Range :RT -30℃ to 100℃
Temperature Accuracy:≤ ± 0.3℃
Dry Bath Incubator ADDB-516
Temperature Range:RT. 5℃ to 160℃
Temperature Accuracy:± 0.5℃ at 40℃ , ± 1℃ at 120℃
Temperature Uniformity:1
Dry Bath Incubator ADDB-513
Temperature Range:RT-25℃ to 100℃
Temperature Accuracy:±0.3℃
Temperature Set Range:-10℃ to 100℃
Dry Bath Incubator ADDB-511
Temperature Range:RT + 5℃ to 150℃
Heating Up Rate:≤ 10 min, 20℃ to 100℃ , ≤ 30 min, 20℃ to 150℃
Temperature Stability :±0.3℃ at 40℃ , ±0.5℃ at 100℃ , ±1℃ at 120℃
Dry Bath Incubator ADDB-519
Temperature Setting Range:0℃ to 100℃
Temperature Control Range:RT +5℃ to 100℃
Temperature Accuracy:≤ ± 0.3℃ at 37℃
Dry Bath Incubator ADDB-509
Temperature Range:RT-30℃ to 100℃
Temperature Accuracy:±0.3°C
Temperature Set Range:-10℃ to 100℃
Dry Bath Incubator ADDB-512
Temperature Range:RT+5℃ to 100℃
Temperature Accuracy:±0.3℃
Temperature Set Range:0℃ to 100℃